Measurement stand for study properties of isolation materials using polarization method
Date
2014Publisher
xmlui.dri2xhtml.METS-1.0.item-identifier-udc
621.317.333Bibliographic entry
Measurement stand for study properties of isolation materials using polarization method / P. Zukowski, T. Koltunowicz, Cz. Kozak, K. Kierczynski, P. Rogalski // Приборы и методы измерений. – 2014. – № 1 (8). – С. 5 - 9.
Abstract
The article presents measurement stand, which is used to measure polarization and depolarization currents, using PDC method, in wet transformer oil-impregnated cellulose isolation. Maximum resolution of measured currents reaches 10-15 A. The accuracy of thermal stabilization, in three-electrode capacitor system, comes to 0,001 °C.